A X 7 3 8 7INSTRUCTIONSIX2-ARCEVAARC ILLUMINATION TIRF SETThis instruction manual is for the Olympus Arc Illumination TIRF Set. To ensure the safety,
74-position excitation filter sliderND filter insertion slotsFilter diameter: 32 mm.Thickness: 1 mm or less.Slider knobLight leak prevention plateExci
IX2-ARCEVA8ASSEMBLYThe filters and sliders are very hot during and immediately after use. Confirm that they are cooled down enoughbefore replacing the
9}When the IX2-RFACEVA mirror unit cassette is used, the operation in step1 below is not necessary because no excitation filter is provided with it.1.
IX2-ARCEVA104. Write the name (or abbreviation) of the mounted filter on one of theprovided handwritten indication stickers and attach it to one of th
114Mounting the IX2-RFAL5Mounting the UV Cut Plate}Refer to the instruction manual for the IX2 reflected fluorescence system.6Mounting the ND Filter}T
IX2-ARCEVA12OPERATION4-1 Preparation1Centering the Burner (Fig. 5)}The following settings are the original settings required for this system.The detai
13Fig. 74. Push in the slit selector lever ² (IN position) and set the wedge prismslider ³ to the intermediate click position to engage the slit and w
IX2-ARCEVA144-2 Total Reflection Fluorescence Observation · The only objective that can be used in total reflection fluorescence observation is the Pl
15Under certain conditions, performance of the system may be adversely affected by factors other than defects. If problemsoccur, please review the fol
Printed on 100% recycled paper with soy ink.
Shinjuku Monolith, 3-1, Nishi Shinjuku 2-chome, Shinjuku-ku, Tokyo, Japan3500 Corporate Parkway, P.O. Box 610, Center Valley, PA 18034-0610, U.S.A.One
IX2-ARCEVACONTENTSCorrect assembly and adjustments are critical for the system to manifest its full performance. If you are going to assemblethe syste
1IMPORTANTThis system has been designed to enable total reflection fluorescence observation using an arc light source(mercury burner or xenon burner).
IX2-ARCEVA21Getting Ready1. This manual pertains only to the handling of the total reflection fluorescence illumination system. Please also read thein
3 . The aperture iris diaphragm for use in ordinary fluorescence observation is not provided. If it is required to use the apertureiris diaphragm, re
IX2-ARCEVA42Maintenance and Storage1. Clean all glass components by wiping gently with gauze. To remove fingerprints or oil smudges, wipe with gauze s
5NOMENCLATURERecommended stage IX-SVL-2Features high operability of irisdiaphragm unit.UV cut plateObjective(Note)Only the PlanApoN60XOTIRFMand PlanAp
IX2-ARCEVA6CONTROLSL-shaped reflected light illuminatorIX2-RFAL(Note)The two filter sliders and apertureiris diaphragm unit are not usedwith this syst